Table 2.
Calculated results for SdNR, FHWM of the ASF, and MTF. Data are separated into the five groups of configurations that were outlined in Sec. 2B. The configuration with 29 projection views, a 28° angular span, and an even dose distribution resulted in the highest “quality factor” value for an exposure of 100 mAs. MMOC stands for more mAs on central projections. LMOC stands for less mAs on central projections.
Group | Number of projections | Total angular span | Entrance dose (mR) | Detector resolution (μm) | mAs distribution | SdNR | FWHM of ASF | MTF at 10% | QF |
---|---|---|---|---|---|---|---|---|---|
1 | 15 | 14° | 727 | 140 | Uniform | 5.72 ± 0.21 | 7.80 | 4.14 ± 0.01 | 0.85 |
1 | 15 | 28° | 682 | 140 | Uniform | 5.44 ± 0.20 | 4.08 | 4.20 ± 0.03 | 1.00 |
2 | 15 | 28° | 682 | 140 | Uniform | 5.44 ± 0.20 | 4.08 | 4.20 ± 0.03 | 1.00 |
2 | 29 | 28° | 656 | 140 | Uniform | 5.81 ± 0.16 | 4.10 | 4.25 ± 0.02 | 1.02 |
3 | 29 | 28° | 656 | 140 | Uniform | 5.81 ± 0.16 | 4.10 | 4.25 ± 0.02 | 1.02 |
3 | 29 | 28° | 665 | 140 | LMOC | 4.97 ± 0.20 | 4.03 | 4.23±0.01 | 0.98 |
3 | 29 | 28° | 675 | 140 | MMOC | 5.04 ± 0.16 | 4.05 | 4.25 ± 0.01 | 0.98 |
4 | 29 | 28° | 385 | 140 | Uniform | 4.32 ± 0.14 | 4.14 | 4.30 ± 0.01 | 0.93 |
4 | 29 | 28° | 523 | 140 | Uniform | 4.87 ± 0.12 | 3.93 | 4.28 ± 0.02 | 0.98 |
4 | 29 | 28° | 656 | 140 | Uniform | 5.81 ± 0.16 | 4.10 | 4.25 ± 0.02 | 1.02 |
4 | 29 | 28° | 791 | 140 | Uniform | 6.06 ± 0.24 | 4.04 | 4.23 ± 0.02 | 1.04 |
5 | 15 | 28° | 682 | 70 | Uniform | 2.97 ± 0.08 | 4.30 | 5.15 ± 0.05 | 0.91 |
5 | 15 | 28° | 682 | 140 | Uniform | 5.44 ± 0.20 | 4.08 | 4.20 ± 0.03 | 1.00 |