Skip to main content
. Author manuscript; available in PMC: 2014 Jun 1.
Published in final edited form as: Magn Reson Imaging. 2012 Dec 7;31(5):783–788. doi: 10.1016/j.mri.2012.11.005

Table 2.

3.0T in-vivo measurements of maximal temperature increases measured at the distal and proximal DBS electrodes by Probe 1 (p1) and Probe 2 (p2), respectively.

Run Number IPG status Geometry Maximal temperature increases (°C)
4-echo fMRI IR SPGR

p1 p2 p1 p2 p1 p2
1 off leads in coil center 1.58 1.55 0.21 0.23 0.24 0.21
2 off leads in coil center 1.54 1.58 0.23 0.21 0.24 0.23
3 on leads in coil center 2.0 2.0 0.29 0.22 0.25 0.23
4 on head shifted 2 cm along z 1.78 1.82 0.21 0.29 0.24 0.31
5 on head rotated by 3° 2.24 2.27 0.44 0.35 0.46 0.35