Table 4.

Phantom experiment. Maximal temperature increases, Δ TMax, measured on DBS electrodes during 3.0T and 1.5T scanning of the ASTM head-and-torso phantom. SARC –average coil SAR displayed by the scanner after patient weight corresponding to auto-prescan TG was entered, SARPM – 6min. average displayed by the power monitor (3.0T).

Pulse Sequence 3.0T 1.5T

Δ TMaxC) Δ TMaxC)
4-echo 2.04 (p1)
1.54 (p2)
0.90 (p1)
0.81 (p2)
fMRI 0.16 (p1)
0.11 (p2)
0.12 (p1)
0.09 (p2)
IR-FSPGR 0.37 (p1)
0.24 (p2)
0.20 (p1)
0.18 (p2)