Table 4.
Application of preferred settings | Condition change | Mascot score | Sequence coverage | Matched spectra | Peptides identified | Spectra/peptide identification |
---|---|---|---|---|---|---|
S.P.P. 50 mm/20 min | DX = 15 s | 2749 | 83% | 248 | 83 | 3.0 |
S.P.P. 50 mm/20 min | DX = 30 s | 2845 | 83% | 221 | 89 | 2.5 |
S.P.P. 50 mm/20 min | DX = 60 s | 1699 | 82% | 152 | 74 | 2.1 |
S.P.P. 50 mm/20 min | DX = 120 s | 1385 | 72% | 114 | 64 | 1.8 |
Data were obtained using the optimum DDA settings determined in Table 3. Specifically, minimum MS signal 1000, repeat count one, and repeat duration 10 s were applied.