Skip to main content
. Author manuscript; available in PMC: 2013 Oct 5.
Published in final edited form as: J Chromatogr A. 2012 Aug 17;1258:76–83. doi: 10.1016/j.chroma.2012.08.036

Table 2.

Effect of sample mass on column efficiency loss.

Particle type Surface area, m2/g Sample mass for 30% loss in plate number, μga
5 μm Halo Fused-Core, 0.6 μm shell 90 18
4 μm Halo Fused-Core, 0.6 μm shell 98 17
4 μm Halo Fused-Core, 0.2 μm shell 30 10
2.7 μm Halo Fused-Core, 0.5 μm shell 135 20
5 μm Totally porous B 170 30
5 μm Totally porous C 450 43
a

Columns: 4.6 mm × 150 mm; sample solute: 1-CI-4-nitrobenzene.