Skip to main content
. 2013 Mar 26;8(3):e59573. doi: 10.1371/journal.pone.0059573

Table 1. Comparison of the new tSEM versus TEM systems.

tSEM TEM
Platform FE-SEMa TEM
Specimen holder multiple grids at a time usually one grid at a time (some special holders)
Illumination rastered probe entire image field
Accelerating voltage 28 kV 120 kV
Detection method transmitted electron detector CCD camera
Automation Routine: Stage translation, image optimization (focus, brightness, etc.), image acquisition, and mosaic. Specialized: Some modules for focus and astigmatism compensation, some mosaic functions.b
Usual operator involvement Identify center of image fields & set scan rotation on each section Repeated specimen exchange & stage translation, physical specimen rotation, image optimization & acquisition
Relative cost of instrument: ∼Half that of manual TEM
(a) Optimal pixel size (in x–y) for analysis 2 nm 2 nm
(b) Section thickness 45 nm 45 nm
Voxel size ( = a×a×b) 2 nm×2 nm×45 nm 2 nm×2 nm×45 nm
(c) Typical number of serial sections 200 200
(d) Maximum single field dimensions 32,768 pixels per side 4,080 pixels per side
(e) Maximum single image field dimensions ( = a×d) 65.54 µm per side 8.160 µm per side
(f) Maximum single image field area ( = e×e) 4,295 µm2 66.59 µm2
(g) Volume of imaged tissue ( = b×c×f) 38,660 µm3 599.3 µm3
Total number of voxels ( = c×d×d) 2.147×1011 voxels 3.329×109 voxels
(h) Operator time to acquire single field image over 200 sections ∼4 hr (plus additional ∼74 hr for unattended image acquisition).c ∼40 hr.
Volume of imaged tissue per operator time ( = g/h) 9664 µm3/hr 14.98 µm3/hr
a

At the time of this writing, Zeiss is the sole source of the FE-SEM system designed for high-resolution large-field transmission imaging of biological samples as described here.

b

TEM can be retrofitted to acquire mosaic images automatically with open source software such as Leginon (National Resource for Automated Molecular Microscopy, The Scripps Research Institute; http://www.leginon.org/) and SerialEM (The Boulder Lab for 3D Electron Microscopy, University of Colorado; http://bio3d.colorado.edu/SerialEM/).

c

Note that the operator time for tSEM remains constant for the same number of serial sections, even when the volume size is increased by montaging.