Table 1. Comparison of the new tSEM versus TEM systems.
tSEM | TEM | ||
Platform | FE-SEMa | TEM | |
Specimen holder | multiple grids at a time | usually one grid at a time (some special holders) | |
Illumination | rastered probe | entire image field | |
Accelerating voltage | 28 kV | 120 kV | |
Detection method | transmitted electron detector | CCD camera | |
Automation | Routine: Stage translation, image optimization (focus, brightness, etc.), image acquisition, and mosaic. | Specialized: Some modules for focus and astigmatism compensation, some mosaic functions.b | |
Usual operator involvement | Identify center of image fields & set scan rotation on each section | Repeated specimen exchange & stage translation, physical specimen rotation, image optimization & acquisition | |
Relative cost of instrument: | ∼Half that of manual TEM | ||
(a) | Optimal pixel size (in x–y) for analysis | 2 nm | 2 nm |
(b) | Section thickness | 45 nm | 45 nm |
Voxel size ( = a×a×b) | 2 nm×2 nm×45 nm | 2 nm×2 nm×45 nm | |
(c) | Typical number of serial sections | 200 | 200 |
(d) | Maximum single field dimensions | 32,768 pixels per side | 4,080 pixels per side |
(e) | Maximum single image field dimensions ( = a×d) | 65.54 µm per side | 8.160 µm per side |
(f) | Maximum single image field area ( = e×e) | 4,295 µm2 | 66.59 µm2 |
(g) | Volume of imaged tissue ( = b×c×f) | 38,660 µm3 | 599.3 µm3 |
Total number of voxels ( = c×d×d) | 2.147×1011 voxels | 3.329×109 voxels | |
(h) | Operator time to acquire single field image over 200 sections | ∼4 hr (plus additional ∼74 hr for unattended image acquisition).c | ∼40 hr. |
Volume of imaged tissue per operator time ( = g/h) | 9664 µm3/hr | 14.98 µm3/hr |
At the time of this writing, Zeiss is the sole source of the FE-SEM system designed for high-resolution large-field transmission imaging of biological samples as described here.
TEM can be retrofitted to acquire mosaic images automatically with open source software such as Leginon (National Resource for Automated Molecular Microscopy, The Scripps Research Institute; http://www.leginon.org/) and SerialEM (The Boulder Lab for 3D Electron Microscopy, University of Colorado; http://bio3d.colorado.edu/SerialEM/).
Note that the operator time for tSEM remains constant for the same number of serial sections, even when the volume size is increased by montaging.