Table 4.

Typical regression equations for calibration curves, retention times, LLOD and LLOQ.

Analyte Slope Regression Intercept r2 RT (minutes) LLOD (μM) LLOQ (μM)
T1AM 0.31 0.006 0.998 7.63 ± 0.05 0.05 0.1
T0AM 0.496 −0.004 0.998 4.95 ± 0.03 0.05 0.1
S-T1AM 0.0204 0.0003 0.989 7.28 ± 0.04 0.1 0.25
Ac-T1AM 0.305 −0.019 0.964 12.86 ± 0.05 0.05 0.1
TA1 0.407 0.016 0.994 13.84 ± 0.11 0.05 0.25
TA0 0.351 0.001 0.999 10.71 ± 0.07 0.05 0.25

RT, retention time, +/− standard deviation.