Skip to main content
. 2013 Mar 11;40(4):041901. doi: 10.1118/1.4793724

Figure 8.

Figure 8

Projection schedule for RT 4D CBCT with displacement bin sorting over one respiratory cycle indicated with dashed vertical lines. Cross marks on the respiratory signal (sine wave) show how only one projection will be sorted into displacement bins 1 through 10. One projection per bin per cycle guarantees no clustering.