Table 3.
Consistency check of dosimeters for output measurement of Co-60.
Dosimeter | Values Obtained | Permissible |
---|---|---|
Ionex-2500/3 | +1.48% to – 0.692% | ± 1.5% |
Farmer- 2570 | +0.974% to – 1.5% | ± 1.5% |
Consistency check of dosimeters for output measurement of Co-60.
Dosimeter | Values Obtained | Permissible |
---|---|---|
Ionex-2500/3 | +1.48% to – 0.692% | ± 1.5% |
Farmer- 2570 | +0.974% to – 1.5% | ± 1.5% |