Skip to main content
. 2008 Dec;4(4):310–318.

Table 1.

Characteristics of the calibration curves of SWAdCS voltammetric determination of pioglitazone HCl

tacc(s) Linearity range (M) Regression equation ip (μA) = a + b C Correlation coefficient (r) LOD (M) LOQ (M)

0 3 × 10-5 - 2 × 10-4 ip = 5.46 + 1.89 C 0.9959 9.92 × 10-6 3.01 × 10-5
30 5 × 10-6 - 9 × 10-5 ip = 0.55 + 6.51 C 0.9983 1.04 × 10-6 3.16 × 10-6
60 2 × 10-6 - 4 × 10-5 ip = 1.64 + 6.53 C 0.9968 6.37 ×10-7 1.93 × 10-6
100 7 × 10-7 - 8 × 10-6 ip = -0.15 + 5.05 C 0.9981 1.60 ×10-7 4.86 × 10-7
150 2 × 10-7 - 3 × 10-6 ip = -0.02 + 7.54 C 0.9967 5.82 ×10-8 1.76 × 10-7
300 4 × 10-8 - 6 × 10-7 ip = 5.56 + 3.23 C 0.9957 8.08 ×10-9 2.45 × 10-8