Skip to main content
. 2009 Mar;5(1):37–43.

Table 2.

Accuracy and precision of the HPLC method for simultaneous determination of ceftriaxone sodium and sulbactam sodium

Drugs Accuracy
Precision
Mean recovery ± SD RSD (%, n=9) 75% 100% 125% label claim

Ceftriaxone 100.88 ± 0.31 0.53 0.34 0.29 0.31
Sulbactam 100.54 ± 0.53 0.31 0.39 0.15 0.44

The chromatographic conditions used were: ODS Hypersil C-18 Column, mobile phase of acetonitrile and 0.005N TBAH (30:70), flow rate of 1.5 ml/min, detection wavelength of 220 nm, room temperature 25°C.