Table 2. Influence of temperature and blood flow on MFPF PO2 measurements: Linear regression model.

Non-standardized coefficients 95.0% CI
Independent variable Regression coefficient Standard error P-value lower limit upper limit
Temperature (base 20 °C)
Constant −2.335 10.483 0.825 −23.744 19.074
PO2-magnitude [(MFPF + CTE)/2] 0.062 0.027 0.031 0.006 0.118
30°C 12.854 11.466 0.271 −10.563 36.271
40°C 19.635 11.392 0.095 −3.629 42.9
Blood flow (base 0.8 L min−1)
Constant −4.707 2.401 0.058 −9.586 0.171
PO2-magnitude [(MFPF + CTE)/2] 0.045 0.004 <0.001 0.037 0.054
1.6 L min−1 −4.461 2.441 0.076 −9.422 0.5
2.4 L min−1 −2.822 2.262 0.221 −7.42 1.776
3.2 L min−1 −2.093 2.328 0.375 −6.825 2.638
4.0 L min−1 −5.188 2.171 0.023 −9.601 −0.776

Linear regression model: magnitude of PO2 measurements, temperature and blood flow as independent variables, differences of PO2 measurements as dependent variable. PO2  =  oxygen partial pressure; MFPF  =  Multi Frequency Phase Fluorimetry; CTE  =  Clark-type electrode; CI  =  confidence interval.