. 2013 May;54(5):1505–1511. doi: 10.1194/jlr.D034553

TABLE 2.

Results for the method validation of F2-isoPs

LLOD (pg inj.) LLOQ (pg inj.) Recoverya (%) Intradayd Interdaye
Precisionb (% CV) Accuracyc (%) Precision (% CV) Accuracy (%)
8-iso-15(R)-PGF 9.1 13.8 62.6 3.1 −3.9 4.7 −5.8
8-iso-PGF 10.6 16.1 71.1 4.6 −21.6 5.3 −22.6
15(R)-PGF 11.0 16.6 73.3 2.9 1.6 4.1 0.3
iPF-IV 7.5 11.3 59.1 2.6 −2.6 3.4 −4.6
iPF-VI 9.2 13.9 58.4 5.2 1.5 8.2 2.3
5-iPF-VI 8.2 12.4 57.9 3.4 1.8 6.5 0.1
(±)5-8,12-iso-iPF-VI 13.8 21.0 71.2 4.1 9.2 8.1 4.1

LLOD, lower limit of detection; injected (inj.), .

a

Recovery was expressed as the mean value of data obtained from plasma samples spiked with 10 μl of working solutions containing 7 ng/ml, or 20 ng/ml of each analyte.

b

Precision is expressed as the mean value of data obtained from plasma samples spiked with 10 μl of working solutions containing 0 ng/ml, 7 ng/ml, or 20 ng/ml of each analyte.

c

Accuracy is expressed as the mean value of data obtained from plasma samples spiked with the 7 or 20 ng/ml working solutions.

d

n = 4 per concentration.

e

n = 12 per concentration.