Table 5.

Infections associated with flexible bronchoscopy

Reference Microorganism(s) No. of contaminated patients after endoscopy No. of infected patients Infection(s) Detection of endoscope contamination Cause(s) of contaminationa
73 P. aeruginosa 11 1 Pneumonia Yes Inappropriate cleaning and disinfection (povidone-iodine); drying with no ethanol flushing
70 P. aeruginosa 8 0 No Yes Lack of manual cleaning; drying with no ethanol flushing
102 P. aeruginosa 35 No data No data Yes Contaminated AER
104 P. aeruginosa 18 3 Pneumonia No data Inadequate cleaning and disinfection (incorrect connectors)
26 P. aeruginosa 8 8 Sepsis, pneumonia Yes Contaminated AER (biofilm on internal plumbing); no protocol for manual cleaning; lack of drying procedure
103 P. aeruginosa 18 3 Pneumonia Yes Contaminated AER; incorrect AER connectors (obstruction of disinfectant flow)
107 P. aeruginosa, S. marcescens 20 (P. aeruginosa), 6 (S. marcescens) 1 Pneumonia Yes Loose biopsy port cap of the contaminated bronchoscopes
30 P. aeruginosa, S. marcescens 41 0 No Yes Insufficient disinfectant exposure; rinsing with tap water; drying with no ethanol flushing
93 P. aeruginosa 32 32 Pneumonia, bacteremia/sepsis, bronchitis, sinusitis Yes Loose biopsy port cap of the contaminated bronchoscopes
105 P. aeruginosa 16 4 Pneumonia Yes Defective biopsy forceps and damaged endoscope channel
64 P. aeruginosa 10 7 Pneumonia, bronchitis Not tested Inappropriate disinfection (povidone-iodine); lack of drying procedure
74 P. aeruginosa 7 7 Pneumonia, bronchitis Yes Contaminated AER (detergent tank); inadequate manual cleaning; drying with no ethanol flushing
106 P. aeruginosa 12 12 Pneumonia, sepsis Yes Multiple manufacturing defects of the contaminated bronchoscope
27 M. tuberculosis 1 0 No No data Inappropriate disinfection (iodophor solution); lack of manual cleaning; drying procedure not described
109 M. tuberculosis 2 1 Lung tuberculosis Yes Inappropriate disinfection (iodophor solution)
111 M. tuberculosis 3 1 Lung tuberculosis Yes Contaminated suction valves (failure to disinfect with glutaraldehyde)
63 M. tuberculosis 3 1 Multidrug-resistant lung tuberculosis Yes Inadequate cleaning and disinfection (no immersion in disinfectant); drying with no ethanol flushing
66 M. tuberculosis 1 0 No No data Inappropriate cleaning and disinfection; lack of drying procedure
108 M. tuberculosis 1 1 Lung tuberculosis No data Inappropriate cleaning and disinfection; drying procedure not described
125 M. tuberculosis 8 0 No Yes Contaminated automated aspiration adaptor (no manual cleaning and disinfection)
104 M. tuberculosis 4 0 No No Inadequate cleaning and disinfection (biopsy port cap was not replaced before loading into the AER)
110 M. tuberculosis 9 3 Lung tuberculosis No Hole in the sheath of the bronchoscope (no leak testing performed)
120 M. tuberculosis 2 0 No No Inadequate cleaning and disinfection between uses in patients (incompatibility of the bronchoscope and AER)
67 M. chelonae 7 0 No No Contaminated AER; drying with no ethanol flushing
124 M. chelonae 72 2 No data Yes Contaminated suction channel (punctured channel)
71 M. chelonae 7 0 No Yes Rinsing with contaminated tap water; drying with no ethanol flushing
126 M. chelonae 4 0 No Yes Contaminated suction channel valve; inadequate endoscope reprocessing
52 M. chelonae 14 0 No No data Contaminated AER; inappropriate disinfection; rinsing with tap water; lack of drying procedure
24 M. chelonae 14 0 No No data Contaminated AER; inappropriate cleaning and disinfection; drying with no ethanol flushing
68 M. chelonae, M. gordonae 14 0 No No data Contaminated AER (rinse water tank); drying with no ethanol flushing
122 M. chelonae 12 0 No Yes Contaminated AER; inadequate endoscope reprocessing
123 M. chelonae 13 0 No No data Contaminated AER; inadequate endoscope reprocessing
76 M. chelonae 15 0 No Yes Contaminated AER; inappropriate disinfection; drying with no ethanol flushing
127 M. chelonae 18 0 No Yes Contaminated suction channels; inappropriate disinfection and drying
25 M. chelonae, Methylobacterium mesophilicum 20 0 No Yes Contaminated AER (presence of biofilm)
111 M. avium complex 2 0 No Yes Contaminated suction valve (failure to disinfect with glutaraldehyde)
104 M. avium complex 7 0 No No Inadequate cleaning and disinfection (incorrect connectors)
56 M. xenopi 2 0 No No data Rinsing with contaminated tap water
51 M. xenopi 20 3 Pulmonary infection, abscess Yes Rinsing with contaminated tap water; inadequate disinfection and drying between uses in patients
121 Mycobacterium spp. (patients), M. chelonae and M. fortuitum (AER) 15 0 No No data Contaminated AER; endoscope reprocessing not described
142 S. marcescens 3 3 Pneumonia Yes Inappropriate cleaning and disinfection (ethanol)
55 S. marcescens 4 0 No Yes Inappropriate disinfection (ethanol); rinsing with tap water
57 S. marcescens 6 5 Bacteremia/sepsis, pneumonia, wound infection Yes Inappropriate cleaning and disinfection; rinsing with tap water; lack of drying procedure
72 S. marcescens 53 (no data about endoscopy) Yes Bacteremia/sepsis, pneumonia, wound infection Yes Drying with no ethanol flushing
153 R. rubra 30 0 No No Contaminated cleaning brushes
69 R. rubra 11 0 No Yes Contaminated suction channel; drying with no ethanol flushing; disinfection procedure not described
156 Bacillus sp. 10 0 No Yes Contaminated suction valves (no disassembling prior to cleaning); inappropriate cleaning and disinfection (ethanol)
54 M. mesophilicum 7 0 No No Rinsing with contaminated tap water; endoscope reprocessing not described
154 Blastomyces dermatitidis 2 0 No Not tested Inadequate manual cleaning; drying procedure not described
155 Legionella pneumophila 3 0 No Not tested Rinsing with contaminated tap water (inadequate maintenance of the filters)
143 Klebsiella pneumoniae, Proteus mirabilis, Morganella morganii, Proteus vulgaris 117 0 No Yes Loose biopsy port cap of the contaminated bronchoscope
a

AER, automated endoscope reprocessor.