Mass Analyzer Mass Resolving Power (mm) m/z Range Throughput (pixels/sec) Tandem MS Capabilities Application
Linear TOF 103 0 – 106 2 – 5 No Peptides/ Proteins
Reflectron TOF 103 – 104 0 – 105 2 – 5 MS2 (post-source decay) Drug/Metabolites/Peptides/Lipids
TOF/TOF 103 – 104 0 – 104 2 – 5 MS2
Q-TOF 103 – 104 100 – 6k < 1 MS2
Linear Ion Trap 102 – 103 50 – 3k < 1 MSn
Triple Quadrupole 102 – 103 50 – 5k > 100 MS2
FT-Orbitrap 104 – 105 50 – 10k < 1 MSn
FT-ICR 104 – 106 100 – 10k 1 – 3 MSn