Skip to main content
. Author manuscript; available in PMC: 2013 Apr 12.
Published in final edited form as: Sci Transl Med. 2012 Sep 19;4(152):152ra129. doi: 10.1126/scitranslmed.3003981

Table 3.

Results from finger-stick testing in 10 healthy volunteers comparing the paper-based test with an automated point of care system (Cholestech LDX).

Volunteer Paper test
(U/l)
Cholestech LDX
(U/l)
ALT AST ALT AST
1 40 40 22 29
2 40 40 22 25
3 40 40 23 26
4 40 40 41 35
5 20 40 27 36
6 20 40 18 23
7 20 40 18 17
8 40 40 45 35
9 20 40 17 18
10 20 40 29 25