. 2012 Apr 1;1(4):56–63. doi: 10.1302/2046-3758.14.2000047

Table III.

Details of the Articuleze tapers in which a taper engagement level (TEL) was easily identified versus those in which there was no such finding (and were indistinguishable from a sterile component)

Characteristics (median and range) TEL identified (n = 25) No TEL identified (n = 23) p-value*
Duration in vivo (mths) 42 (12 to 74) 43 (12 to 75) 0.961
Taper angle (°)   5.639 (5.584 to 5.685)   5.629 (5.557 to 5.672) 0.471
Diametral clearance (µm) 88.8 (70.2 to 112.8) 89.6 (70.2 to 119.4) 0.855
Horizontal lever arm distance (mm) 17.94 (10.26 to 24.53) 16.04 (12.34 to 20.89) 0.007
Surface wear (mm3/yr)   2.03 (0.51 to 6.29)   3.57 (0.51 to 13.67) 0.069

* Mann-Whitney U test