Abstract
Biofilms have been observed in the fluid pathways of hemodialysis machines. The impacts of four biocides used for the disinfection of hemodialysis systems were tested against Candida parapsilosis sensu stricto and Candida orthopsilosis biofilms generated by isolates obtained from a hydraulic circuit that were collected in a hemodialysis unit. Acetic acid was shown to be the most effective agent against Candida biofilms. Strategies for effective disinfection procedures used for hemodialysis systems should also seek to kill and inhibit biofilms.
TEXT
The water treatment system is a matter of major concern in hemodialysis, and reports have described its contamination by biofilms (1, 2). For disinfection, active chemical agents (biocides) have been introduced into routine practice (3). Hemodialyzers are disinfected with peracetic acid, and hemodialysis systems (which include hemodialysis machines, the water supply, water treatment systems, and distribution systems) are typically disinfected using chlorine-based disinfectants at an aqueous concentration of 500 ppm (3–5). Low-pH cleaning agents have also been used as disinfectants (4), and 3% hydrogen peroxide (vol/vol) may be used to treat biofilms on implants, on the implant-surrounding tissue, on the skin surface, or on infected wounds without devices (6).
In South American hospitals, the incidence of Candida parapsilosis is greater than that of Candida albicans (7, 8), and previous results from our laboratory (9–11) have shown that C. parapsilosis is predominantly implicated in the contamination of water samples collected at a hemodialysis center. Recently, C. parapsilosis isolates were classified into three distinct species: C. parapsilosis sensu stricto, Candida orthopsilosis, and Candida metapsilosis (12).
Although the nephrology community has been provided with information about the effects of disinfectants on microbial suspensions (3–5, 13, 14), few studies have evaluated the efficacy of disinfection in the presence of biofilms, particularly fungal biofilms. Therefore, the biocidal efficacy of the commercially available concentrations of biocides used for the disinfection of hemodialysis systems was evaluated against both C. parapsilosis sensu stricto and C. orthopsilosis biofilms. Additionally, the effects of concentration on the time required for effective treatment were compared between the more efficacious biocides and the standard established by legislation.
One hundred C. parapsilosis isolates were obtained from the sampling of water held in the hemodialysis unit studied (disinfection with a 30-min exposure to sodium hypochlorite at 500 ppm on a daily basis) located in the state of São Paulo, Brazil, between March 2006 and March 2007. The identities of the yeast isolates were determined by using a PCR-based method (12, 15) and specific primers directed against the secondary alcohol dehydrogenase (SADH) gene; 53 were assigned to the species C. parapsilosis sensu stricto, and 47 to C. orthopsilosis. The strains were stored in sterile, distilled water (16) in our laboratory. Fifteen strains each of two Candida species, namely, C. parapsilosis sensu stricto (WCP3, WCP4, WCP6, WCP8, WCP10, WCP14, WCP16, WCP17, WCP24, WCP82, WCP83, WCP87, WCP88, WCP104, WCP108) and C. orthopsilosis (WCO33, WCO45, WCO52, WCO53, WCO125, WCO139, WCO147, WCO154, HMCOB, HMCOC, HMCOM, HMCOQ, HMCOQ1, HMCOR, HMCOU), were selected for this study. All Candida strains were subcultured on Sabouraud's dextrose agar (SDA; Gibco Ltd., Paisley, United Kingdom) and maintained at 4°C during the experimental period.
Acetic acid (C2H4O2; 0.5% [vol/vol]; Merck, Darmstadt, Germany), hydrogen peroxide (H2O2; 3% [vol/vol]; Synth, SP, Brazil), sodium hypochlorite (NaOCl; 2% [wt/vol]; Merck), and a commercial biocide made of peracetic acid and hydrogen peroxide (CH3CO3H and H2O2; 1.34% [vol/vol] and 4.2% [wt/vol]; Proxitane; Fresenius Medical Care, Bad Homburg, Germany) were tested against Candida biofilms. All of the biocide stock solutions were diluted in filter-sterilized phosphate-buffered saline (PBS; 10 mM phosphate buffer, 2.7 mM potassium chloride, 137 mM sodium chloride, pH 7.4; Sigma Chemical Co., St. Louis, MO), followed by further dilution in Roswell Park Memorial Institute (RPMI) 1640 medium supplemented with l-glutamine and buffered with 0.165 M morpholinepropanesulfonic acid (MOPS; Sigma) at concentrations ranging from 0.002 to 2.7 g/liter for acetic acid, 0.01 to 15 g/liter for hydrogen peroxide, 0.02 to 20 g/liter for sodium hypochlorite, 0.006 to 6.7 g/liter for peracetic acid, and 0.02 to 21 g/liter for hydrogen peroxide.
All of the studied strains were screened for their ability to form biofilms by following the methodology previously described by Ramage et al. (17), and semiquantitative measurement of the growth biofilms was performed using the 3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyl tetrazolium bromide (MTT; Sigma) reduction assay (18–21) at 540 nm. C. albicans SC5314 was used as the biofilm control strain (22, 23).
The experiments were divided into three set of experiments: (i) inhibition of biofilm formation, (ii) treatment of preformed biofilms with all biocides for 48 h of incubation, (iii) time-kill curves for more active biocides on established Candida biofilms (24 h old). For each group of the assays, five strains of C. orthopsilosis, five strains of Candida parapsilosis sensu stricto, and the strain C. albicans SC 5314 were used. For the inhibition assays, the biocides were added at the time of inoculation. For the treatment assays, the biocides were diluted in RPMI and added to established biofilms (24 h). Microtiter wells containing heat-killed Candida (106 CFU/ml) were included as negative controls, and biofilms incubated in the absence of biocides were included as positive controls (24). After the desired contact times (0.5 min, 1 min, 5 min, 15 min, 30 min, 60 min, and 48 h), the test product was discarded, and the wells were washed twice with PBS. The use of a neutralizer was unnecessary; preliminary studies showed no differences between wells with and without neutralizer (data not shown). After each exposure to a biocide, the percentage of the metabolically active total cell population was determined by CFU enumeration on Sabouraud dextrose agar (Difco Laboratories, Detroit, MI) incubated at 30°C for 48 h. A standard curve was constructed to correlate the results obtained from the MTT assay with the calculated CFU/ml (limit of detection, 10 CFU/ml), and the data were expressed as the decimal logarithm. No change in optical density over the negative control was equal to ≤10 CFU/ml and was defined as the lowest concentration of biocide which resulted in total inhibition or in total killing of sessile cells. The experiments were repeated three times with at least four replicates for each time point.
The groups were compared using a t test and a one-way analysis of variance (ANOVA), followed by the post hoc Tukey test when all of the disinfectants were compared with the standard following 30 min of incubation. Logistic regression analyses were used to investigate relationships between MTT assay absorbance readings and the total viable count using the SPSS 15.0 software package (SPSS Inc., Chicago, IL). All tests were performed using a significance level of P values of ≤0.05.
Overall, biofilm production by Candida was observed for 34 of C. parapsilosis sensu stricto strains and 38 of C. orthopsilosis strains. MTT activity was linearly associated with CFU counts, and the correlation coefficients (R2) were 0.8831, 0.9045, and 0.9178 for C. parapsilosis sensu stricto, C. orthopsilosis, and the reference strain C. albicans SC 5314, respectively. Strains for which the optical density was of 2.84 ± 0.059 and 2.3 ± 0.094 for C. orthopsilosis and C. parapsilosis sensu stricto, respectively, were scored as better biofilm formers and were selected for antibiofilm assays.
Candida biofilms were exposed to the biocides. After the biofilms were treated with NaOCl, C2H4O2, or CH3COOOH and H2O2, the results were similar for all the tested species (Table 1). However, when H2O2 was evaluated, the efficiency for C. orthopsilosis was the lowest compared to that of other Candida species. Still, the biocides concentration needed to inhibit the biofilm formation was dependent of the Candida species (Table 1). Additionally, a standard biocide (NaOCl, 0.5 g/liter) was compared to the experimental biocides with the best results (Fig. 1). A solution of NaOCl had little effect on the biofilm viability of all of the preformed (24-h) Candida biofilms (Fig. 1A to C). C. orthopsilosis and C. parapsilosis sensu stricto biofilm-embedded cells were killed after 1 min of C2H4O2 (0.33 g/liter) contact, whereas for reference strain C. albicans SC 5314 biofilm cells (Fig. 2), the same effect was obtained after 2 min of exposure.
Table 1.
The in vitro efficacy of biocides against Candida orthopsilosis and C. parapsilosis sensu stricto biofilms
| Candida species | Compounda | Range(s) of concn (g/liter) tested | Concn required to kill biofilm cells in treatment assays (g/l)b,c | Concn required to inhibit biofilm formation (g/l)b,d |
|---|---|---|---|---|
| C. orthopsilosis | NaOCl | 0.02–20 | 2.5 ± 1.36 | 1.25 ± 0.68 |
| H2O2 | 0.01–15 | 3.75 ± 1.02 | 0.93 ± 0.09 | |
| C2H4O2 | 0.002–2.7 | 0.33 ± 0.07 | 0.04 ± 0.01 | |
| CH3COOOH and H2O2 | 0.006–6.7 and 0.02–21 | 0.83 ± 0.18 and 2.6 ± 0.58 | 0.21 ± 0.09 and 0.6 ± 0.22 | |
| C. parapsilosis sensu stricto | NaOCl | 0.02–20 | 2.5 ± 0.66 | 2.5 ± 0.55 |
| H2O2 | 0.01–15 | 1.87 ± 0.51 | 0.47 ± 0.03 | |
| C2H4O2 | 0.002–2.7 | 0.33 | 0.04 | |
| CH3COOOH and H2O2 | 0.006–6.7 and 0.02–21 | 0.83 ± 0.29 and 2.6 ± 0.93 | 0.41 ± 0.18 and 0.3 ± 0.13 | |
| C. albicans (SC 5314) | NaOCl | 0.02–20 | 2.5 ± 0.55 | 2.5 ± 1.11 |
| H2O2 | 0.01–15 | 1.87 ± 0.42 | 0.47 ± 0.06 | |
| C2H4O2 | 0.002–2.7 | 0.33 ± 0.09 | 0.02 ± 0.01 | |
| CH3COOOH and H2O2 | 0.006–6.7 and 0.02–21 | 0.83 ± 0.23 and 2.6 ± 0.71 | 0.21 ± 0.04 and 0.6 ± 0.16 |
NaOCl, sodium hypochlorite; H2O2, hydrogen peroxide; C2H4O2, acetic acid; H3COOOH and H2O2, peracetic acid and hydrogen peroxide (Proxitane).
The inoculum used for the inhibition assays was identical to the cell density used for the treatment assays (106 cells/ml). The concentrations obtained from the optical density at 540 nm (MTT assay) correlated with the total viable count (CFU/ml). The values are means ± 1 standard deviation (SD) from four experiments for each disinfectant. The experiments were repeated three times with at least four replicates at each time point.
Strains: C. orthopsilosis, WCO33, WCO45, WCO52, WCO53, WCO125; C. parapsilosis sensu stricto, WCP3, WCP4, WCP6, WCP8, WCP10.
Strains: C. orthopsilosis, HMCOM, HMCOQ, HMCOQ1, HMCOR, HMCOU; C. parapsilosis sensu stricto, WCP83, WCP87, WCP88, WCP104, WCP108.
Fig 1.

A comparison of the changes in the percent survival of C. orthopsilosis (WCO139, WCO147, WCO154, HMCOB, HMCOC) (A), C. parapsilosis (WCP14, WCP16, WCP17, WCP24, WCP82) (B), and C. albicans SC 5314 (C) biofilms after treatment with disinfectants (H2O2 [hydrogen peroxide], 1.87 g/liter or 3.75 g/liter; C2H4O2 [acetic acid], 0.33 g/liter); or the standard biocide (NaOCl [sodium hypochlorite], 0.5 g/liter) obtained by the MTT reduction assay. Biofilms were cultured in RPMI medium for 24 h and tested against biocides for 5, 10, 15, 30, and 60 min of contact. Each point represents the average of four measurements, and the error bars represent the standard deviations. The experiments were repeated three times with at least four replicates at each time point, *, P ≤ 0.00001, comparing all disinfectants with the standard in 30 min.
Fig 2.

The effects of acetic acid on preformed Candida biofilms. The metabolic activities of C. orthopsilosis (HMCOM, HMCOQ, HMCOQ1, HMCOR, HMCOU), C. parapsilosis (WCP83, WCP87, WCP88, WCP104, WCP108), and C. albicans SC 5314 were measured by the MTT reduction assay. Sessile yeast cells were exposed to acetic acid at a concentration of 0.33 g/liter for periods of 30 s to 5 min, and the biofilm reduction was compared with the untreated control. The bars represent the averages of three MTT measurements, and the brackets denote the standard deviations. The experiments were repeated three times with at least four replicates at each time point.
The presence of biofilm in dialysis systems is a point of concern, first because biofilms continuously release microbial components, such as toxin fragments, peptides, and polysaccharides, able to traverse dialysis membranes (14). To our knowledge, this is the first published report of the effects of biocides used for hemodialysis disinfection on C. parapsilosis sensu stricto and C. orthopsilosis biofilms.
In our study using preformed (24-h) Candida biofilms, 48 h of treatment with NaOCl (2.5 g/liter) killed biofilm cells of all Candida biofilms and inhibited biofilm formation by C. parapsilosis sensu stricto and C. albicans SC 5314. For C. orthopsilosis, the concentration of 1.25 g/liter of NaOCl inhibited biofilm formation (Table 1). The tested concentrations are extremely corrosive, irritate mucosal surfaces, and are higher than those recommended by previous legislation (4, 5) for the disinfection of hemodialysis systems (500 ppm; 0.5 g/liter). However, these guidelines are not adequate for sessile microorganisms.
A previous work (25) focusing on microbial contamination in water treated for use in dialysis reported that the impact of treatment with hypochlorite on the water quality was negative due to detachment of the biofilm and resuspension of heavy loads of microorganisms in water immediately after disinfection. Roeder et al. (26), who investigated the effect of disinfectants against biofilms in drinking water, described that intervention eventually selects persistent microorganisms that can live on the waste of dead cells present in the biofilm. Limited disinfection efficacy against biofilm attached to a granular activated carbon filter has been demonstrated by LeChevallier et al. (27). Additionally, Norman et al. (28) have provided evidence that a 165-μm-thick biofilm chlorinated at 2.0 or 4.1 g/day rapidly returned to its original thickness as soon as the treatment was discontinued.
Peracetic acid and hydrogen peroxide also suppressed biofilm formation, and the former inhibited Candida orthopsilosis and C. albicans biofilm formation at concentrations below 0.3%. This combination of disinfectants has been successfully employed in Brazil (29) to reprocess dialyzers from the same patients. The effectiveness of the mixture of these biocides against bacterial biofilms in water treatment systems has also been described in the literature (26, 30, 31). However, in addition to being very expensive, the combination of hydrogen peroxide and peracetic acid can damage the hemodialysis hydraulic system.
Therefore, in an attempt to identify more economical and less harmful products, two compounds, hydrogen peroxide and acetic acid, were tested against Candida biofilms. In this work, H2O2 at concentrations below 3% (Table 1) effectively inhibited biofilm growth and killed the biofilm populations, with the exception of C. orthopsilosis (3.75 g/liter). Nett et al. (32) reported that a higher concentration of H2O2 was needed to reduce the C. parapsilosis biofilm burden. Additionally, H2O2 efficacy in water system disinfection has been described by Wong et al. (33).
Acid sanitizers are generally utilized at a concentration of 100 ppm or 0.1 g/liter (4). In this study, acetic acid proved to be a better antibiofilm agent against all of the assayed strains, a result similar to that obtained for biofilms consisting of Listeria monocytogenes, Staphylococcus aureus, and Salmonella spp. (34–36).
Since no method presently exists that detects biofilm in vivo with sufficient sensitivity to confirm its eradication, efforts in optimizing the cleaning and disinfection procedures used for hemodialysis systems should also seek to achieve biofilm, especially fungal biofilms. As demonstrated by this study, the standard current biocide (sodium hypochlorite, 500 ppm or 0.5 g/liter) failed to destroy the Candida biofilms tested. Our results showed that hydrogen peroxide or acetic acid may be alternatives, because they are more efficient than sodium hypochlorite. Further studies are needed to determine whether these agents can also be effective against additional Candida species. Ensuring water quality results in reduced patient morbidity and number of hospitalizations and improved quality of life.
ACKNOWLEDGMENTS
This work was supported by grants from the Scientific Development Support Program of Faculdade de Ciências Farmacêuticas at UNESP (PADC/FCFAr-UNESP proc. 2009/54), Coordenação de Aperfeiçoamento dePessoal de Nível Superior (CAPES), and FAPESP (2012/01270-9).
Footnotes
Published ahead of print 11 March 2013
REFERENCES
- 1. Cappelli G, Ravera F, Ricardi M, Ballestri M, Perrone S, Albertazzi A. 2005. Water treatment for hemodialysis: a 2005 update. Contrib. Nephrol. 149:42–50 [DOI] [PubMed] [Google Scholar]
- 2. Man NK, Degremont A, Darbord JC, Collet M, Vaillant P. 1998. Evidence of bacterial biofilm in tubing from hydraulic pathway of hemodialysis system. Artif. Organs 22:596–600 [DOI] [PubMed] [Google Scholar]
- 3. Centers for Disease Control and Prevention 2008. Guideline for disinfection and sterilization in healthcare facilities, 2008. CDC, Atlanta, GA: http://www.cdc.gov/hicpac/pdf/guidelines/Disinfection_Nov_2008.pdf [Google Scholar]
- 4. Association for the Advancement of Medical Instrumentation (AAMI) 2006. Water treatment equipment for hemodialysis applications. ANSI/AAMI RD62, Arlington, VA [Google Scholar]
- 5. Brazilian Ministry of Health 15 June 2004, posting date Technical regulations for the operation of dialysis services, 2004. Directives no. 154/MS. Brazilian Ministry of Health, Brasilia, Brazil: (In Portuguese.). http://www.samu.fortaleza.ce.gov.br/legislacao/Dialise_ANVISA_RDC_154.pdf [Google Scholar]
- 6. Block SS. 2001. Peroxygen compounds, p 185–204 In Block SS. (ed), Disinfection, sterilization, and preservation. Lippincott Williams & Wilkins, Philadelphia, PA [Google Scholar]
- 7. Medrano DJ, Brilhante RS, Cordeiro Rde A, Rocha MF, Rabenhorst SH, Sidrim JJ. 2006. Candidemia in a Brazilian hospital: the importance of Candida parapsilosis. Rev. Inst. Med. Trop. São Paulo 48:17–20 [DOI] [PubMed] [Google Scholar]
- 8. Nucci M, Queiroz-Telles F, Tobón AM, Restrepo A, Colombo AL. 2010. Epidemiology of opportunistic fungal infections in Latin America. Clin. Infect. Dis. 51:561–570 [DOI] [PubMed] [Google Scholar]
- 9. Pires RH, Santos JM, Zaia JE, Martins CH, Mendes-Giannini MJ. 2011. Candida parapsilosis complex water isolates from a haemodialysis unit: biofilm production and in vitro evaluation of the use of clinical antifungals. Mem. Inst. Oswaldo Cruz 106:646–654 [DOI] [PubMed] [Google Scholar]
- 10. Pires-Gonçalves RH, Sartori FG, Montanari LB, Zaia JE, Melhem MS, Mendes-Giannini MJ, Martins CH. 2008. Occurrence of fungi in water used at a haemodialysis centre. Lett. Appl. Microbiol. 46:542–547 [DOI] [PubMed] [Google Scholar]
- 11. Varo SD, Martins CH, Cardoso MJ, Sartori FG, Montanari LB, Pires-Gonçalves RH. 2007. Isolation of filamentous fungi from water used in a hemodialysis unit. Rev. Soc. Bras. Med. Trop. 40:326–331 [DOI] [PubMed] [Google Scholar]
- 12. Tavanti A, Davidson AD, Gow NAR, Maiden MCJ, Odds FC. 2005. Candida orthopsilosis and Candida metapsilosis spp. nov. to replace Candida parapsilosis groups II and III. J. Clin. Microbiol. 43:284–292 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Barraclough KA, Hawley CM, Playford EG, Johnson DW. 2009. Prevention of access-related infection in dialysis. Expert Rev. Anti Infect. Ther. 7:1185–1200 [DOI] [PubMed] [Google Scholar]
- 14. Cappelli G, Riccardi M, Perrone S, Bondi M, Ligabue G, Albertazzi A. 2006. Water treatment and monitor disinfection. Hemodial. Int. 10(Suppl 1):S13–S18 [DOI] [PubMed] [Google Scholar]
- 15. Tavanti A, Hensgens LAM, Ghelardi E, Campa M, Senesi S. 2007. Genotyping of Candida orthopsilosis clinical isolates by amplification fragment length polymorphism reveals genetic diversity among independent isolates and strain maintenance within patients. J. Clin. Microbiol. 45:1455–1462 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. McGinnis MR, Padhye AA, Ajello L. 1974. Storage of stock cultures of filamentous fungi, yeasts, and some aerobic actinomycetes in sterile distilled water. Appl. Microbiol. 28:218–222 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Ramage G, Vande Walle K, Wickes BL, Lopez-Ribot JL. 2001. Standardized method for in vitro antifungal susceptibility testing of Candida albicans biofilms. Antimicrob. Agents Chemother. 45:2475–2479 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Basso FG, Oliveira CF, Fontana A, Kurachi C, Bagnato VS, Spolidório DM, Hebling J, de Souza Costa CA. 2011. In vitro effect of low-level laser therapy on typical oral microbial biofilms. Braz. Dent. J. 22:502–510 [DOI] [PubMed] [Google Scholar]
- 19. Kovaleva J, Degener JE, van der Mei HC. 2010. Mimicking disinfection and drying of biofilms in contaminated endoscopes. J. Hosp. Infect. 76:345–350 [DOI] [PubMed] [Google Scholar]
- 20. Schillaci D, Arizza V, Dayton T, Camarda L, Di Stefano V. 2008. In vitro anti-biofilm activity of Boswellia spp. oleogum resin essential oils. Lett. Appl. Microbiol. 47:433–438 [DOI] [PubMed] [Google Scholar]
- 21. Silveira LC, Charone S, Maia LC, Soares RM, Portela MB. 2009. Biofilm formation by Candida species on silicone surfaces and latex pacifier nipples: an in vitro study. J. Clin. Pediatr. Dent. 33:235–240 [DOI] [PubMed] [Google Scholar]
- 22. Pierce CG, Uppuluri P, Tristan AR, Wormley FL, Jr, Mowat E, Ramage G, Lopez-Ribot JL. 2008. A simple and reproducible 96-well plate-based method for the formation of fungal biofilms and its application to antifungal susceptibility testing. Nat. Protoc. 3:1494–1500 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Ramage G, Bachmann S, Patterson TF, Wickes BL, López-Ribot JL. 2002. Investigation of multidrug efflux pumps in relation to fluconazole resistance in Candida albicans biofilms. Antimicrob. Agents Chemother. 49:973–980 [DOI] [PubMed] [Google Scholar]
- 24. Martinez LR, Casadevall A. 2007. Cryptococcus neoformans biofilm formation depends on surface support and carbon source and reduces fungal cell susceptibility to heat, cold, and UV light. Appl. Environ. Microbiol. 73:4592–4601 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Morin P. 2000. Identification of the bacteriological contamination of a water treatment line used for haemodialysis and its disinfection. J. Hosp. Infect. 45:218–224 [DOI] [PubMed] [Google Scholar]
- 26. Roeder RS, Lenz J, Tarne P, Gebel J, Exner M, Szewzyk U. 2010. Long-term effects of disinfectants on the community composition of drinking water biofilms. Int. J. Hyg. Environ. Health 213:183–189 [DOI] [PubMed] [Google Scholar]
- 27. LeChevallier MW, Cawthon CD, Lee RG. 1988. Inactivation of biofilm bacteria. Appl. Environ. Microbiol. 54:2492–2499 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Norman G, Characklis WG, Bryers JD. 1977. Control of microbial fouling in circular tubes with chlorine. Dev. Ind. Microbiol. 18:581–590 [Google Scholar]
- 29. Mazzola PG, Martins AM, Penna TC. 2006. Chemical resistance of the Gram-negative bacteria to different sanitizers in a water purification system. BMC Infect. Dis. 16:131. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Exner M, Tuschewitzki GJ, Scharnagel J. 1987. Influence of biofilms by chemical disinfectants and mechanical cleaning. Zentralbl. Bakteriol. Mikrobiol. Hyg. 183:549–563 [PubMed] [Google Scholar]
- 31. Monarca S, Garusi G, Gigola P, Spampinato L, Zani C, Sapelli PL. 2002. Decontamination of dental unit waterlines using disinfectants and filters. Minerva Stomatol. 51:451–459 [PubMed] [Google Scholar]
- 32. Nett JE, Guite KM, Ringeisen A, Holoyda KA, Andes DR. 2008. Reduced biocide susceptibility in Candida albicans biofilms. Antimicrob. Agents Chemother. 52:3411–3413 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Wong WC, Dudinsky LA, Garcia VM, Ott CM, Castro VA. 2010. Efficacy of various chemical disinfectants on biofilms formed in spacecraft potable water system components. Biofouling 26:583–586 [DOI] [PubMed] [Google Scholar]
- 34. Akiyama H, Yamasaki O, Tada J, Arata J. 1999. Effects of acetic acid on biofilms formed by Staphylococcus aureus. Arch. Dermatol. Res. 291:570–573 [DOI] [PubMed] [Google Scholar]
- 35. Hasegawa A, Hara-Kudo Y, Kumagai S. 2011. Survival of Salmonella strains differing in their biofilm-formation capability upon exposure to hydrochloric and acetic acid and to high salt. J. Vet. Med. Sci. 73:1163–1168 [DOI] [PubMed] [Google Scholar]
- 36. Rieu A, Guzzo J, Piveteau P. 2010. Sensitivity to acetic acid, ability to colonize abiotic surfaces and virulence potential of Listeria monocytogenes EGD-e after incubation on parsley leaves. J. Appl. Microbiol. 108:560–570 [DOI] [PubMed] [Google Scholar]