Table 2.

Calculated effective doses from the three-dimensional (3D) eXama and Pan eXam Plus 3Da for ICRP 60 and ICRP 103 recommendations of the International Commission on Radiological Protection (ICRP)

Device Measurement ID FOV (cm) Voxel size (mm) Effective dose ICRP 60 (µSv) Effective dose ICRP 103 (µSv) Deviation (%)
KaVo 3D eXam 1 23 × 17 pre scan 1.7 3.1 82.4
2 23 × 17 0.400 56.4 72.0 27.7
3 16 × 13 0.300 62.7 106.7 70.2
4 16 × 8 0.250 67.6 169.8 151.2
5 16 × 8 0.300 18.4 44.5 141.8
6 16 × 4 maxilla 0.125 25.6 67.6 164.1
7 16 × 4 maxilla 0.300 12.3 32.8 166.7
8 16 × 4 mandible 0.125 29.2 76.3 161.3
9 16 × 4 mandible 0.300 14.3 37.7 163.6
10 8 × 8 0.125 48.4 122.1 152.3
11 8 × 8 0.300 22.8 61.6 170.2
KaVo Pan eXam Plus 3D 12 6.4 × 7.8 pre scan 0.4 1.1 175.0
13 6.1 × 4.1 maxilla 0.200 11.7 40.2 243.6
14 6.1 × 4.1 maxilla 0.133 23.8 79.2 232.8
15 6.1 × 4.1 mandible 0.200 20.9 49.3 135.9
16 6.1 × 4.1 mandible 0.133 48.8 114.8 135.2
17 6.4 × 7.8 maxilla 0.300 27.4 79.3 189.4
18 6.4 × 7.8 maxilla 0.200 38.0 124.9 228.7
19 6.4 × 7.8 mandible 0.300 39.3 109.6 178.9
20 6.4 × 7.8 mandible 0.200 68.4 183.7 168.6
a

Manufactured by KaVo Dental, Biberach, Germany.