Table 2. RDT test line specificities for diagnosis of malaria infection in controls and HAT.

Malaria RDT Test line % specificity controlsa % specificity HATb p-valuec
1 Paracheck Pf HRP-2 100 (95.4–100) 98.8 (93.5–100) 1
2 ICT Malaria Pf Cassette Test HRP-2 98.7 (93.1–100) 87.9 (80.0–94.1)d 0.009
3 Advantage Pan Malaria Card pan-pLDH 86.3 (76.2–93.2) 30.0 (20.3–41.3) <0.001
4 Malaria Antigen Pf (HRP-2/pLDH) HRP-2 97.5 (91.1–99.7) 83.3 (73.6–90.6) 0.003
Pf-pLDH 100 (95.4–100) 98.8 (93.5–100) 1
Combined 97.5 (91.1–99.7) 83.3 (73.6–90.6) 0.003
5 SD malaria Ag Pf/Pan HRP-2 100 (95.4–100) 90.5 (82.1–95.8) 0.007
pan-pLDH 100 (95.1–100) 80.0 (69.6–88.1) <0.001
Combined 100 (95.1–100) 76.3 (65.4–85.1) <0.001
6 SD Malaria Antigen Pf Pf-pLDH 97.5 (91.2–99.7) 97.6 (91.6–99.7) 1
pan-pLDH 97.3 (90.5–99.7) 96.3 (89.4–99.2) 1
Combined 97.3 (90.5–99.7) 96.3 (89.4–99.2) 1
7 ICT Malaria Combo Cassette Test HRP-2 87.3 (78.0–93.8) 67.9 (56.8–77.6) 0.005
pan-pLDH 86.3 (76.2–93.2) 13.8 (7.1–23.3) <0.001
Combined 79.5 (68.4–88.0) 11.3 (5.3–20.3) <0.001
8 Carestart Malaria HRP2/pLDH (Pf/Pan) Combo Test HRP-2 92.6 (84.2–97.2) 79.8 (69.6–87.7) 0.03
pan-pLDH 98.6 (92.6–99.8) 76.3 (65.4–85.1) <0.001
Combined 91.8 (83.0–96.9) 61.3 (50.5–72.0) <0.001
9 Carestart Malaria pLDH (Pf/pan) Pf-pLDH 100 (95.4–100) 100% (95.7–100) 1
pan-pLDH 100 (95.1–100) 97.5 (91.3–99.7) 0.5
Combined 100 (95.1–100) 97.5 (91.3–99.7) 0.5
10 First Response Malaria Ag (pLDH/HRP2) Combo Rapid Diagnostic Test HRP-2 98.7 (93.1–100) 95.2 (88.3–98.7) 0.4
pan-pLDH 98.6 (92.6–99.8) 88.8 (79.7–94.7) 0.02
Combined 97.3 (90.5–99.7) 86.3 (76.7–92.9) 0.02
a

HRP-2 and Pf-pLDH N = 79 Pf-PCR negative controls, pan-pLDH test lines n = 73 malaria-PCR negative controls.

b

HRP-2 and Pf-pLDH N = 84/85 Pf-PCR negative HAT patients (1 blood sample missing), pan-pLDH N = 80/81 malaria-PCR negative HAT patients (1 blood sample missing).

c

p values for Fisher exact test.

d

N = 83. The test line for 1 sample could not be scored due to poor background clearance of the test strip.