Skip to main content
. Author manuscript; available in PMC: 2014 Feb 5.
Published in final edited form as: Anal Chem. 2013 Jan 14;85(3):1896–1904. doi: 10.1021/ac303188j

Table 3.

Analytical recovery (bias) and imprecision data for ARVs in meconium

Analytical Recovery (% of target concentration) Imprecision (% RSD, n=20)

Analyte Expected
conc.
(ng/g)
Intra-day, n=4 Inter-day, n=20

Mean Range Mean Range Pooled
Intra-day
Inter-
day
Total
30 107.5 100.7 – 117.0 105.0 86.3 – 117.0 4.7 7.7 9.0
TFV 300 106.2 101.0 – 114.0 107.9 101.0 – 117.0 3.5 3.8 5.2
2000 101.4 98.0 – 105.5 103.8 98.0 – 115.0 4.1 1.0 4.2

150 110.3 106.7 – 114.7 106.4 98.0 – 114.7 3.5 2.0 4.0
3TC 750 107.9 102.0 – 111.3 106.7 100.4 – 112.0 3.7 0.8 3.8
2000 106.5 105.0 – 107.5 103.7 96.0 – 114.5 3.5 3.1 4.7

30 105.7 104.7 – 106.7 103.0 96.3 – 111.0 2.7 1.4 3.0
FTC 300 103.8 101.0 – 108.0 104.5 96.3 – 111.3 3.0 1.4 3.3
2000 98.6 96 – 101.5 99.6 90.5 – 108.0 3.8 3.1 5.0

30 107.8 102.0 – 114.7 111.9 98.7 – 119.0 4.6 1.1 4.7
ABC 300 110.9 108.3 – 116.7 109.2 97.0 – 118.3 3.6 5.3 6.4
2000 96.9 84.0 – 102.5 95.3 84.0 – 104.5 5.4 3.3 6.3

30 120.3 105.7 – 127.7 109.7 88.0 – 127.7 8.1 8.6 11.9
ABC-Ga 300 118.8 104.0 – 124.3 110.8 81.3 – 129.3 12.4 4.1 13.1
2000 114.1 96.5 – 122.0 102.8 72.5 – 137.5 15.5 9.0 17.9

150 120.2 108.7 – 128.0 105.1 70.0 – 128.0 10.2 13.0 16.5
ABC-Ca 750 120.6 112.0 – 133.3 111.8 91.5 – 137.3 11.9 5.8 13.2
2000 111.8 104.5 – 117.0 103.4 73.5 – 138.5 12.8 8.7 15.5

30 108.3 104.3 – 115.3 111.0 104.3 – 116.7 3.6 0.7 3.7
NVP 300 109.4 100.7 – 117.7 109.7 98.3 – 117.7 4.5 1.1 4.7
2000 90.8 86.5 – 95.0 92.0 84.0 – 99.5 2.9 3.1 4.2

30 84.5 80.7 – 90.0 89.3 80.3 – 108.0 4.2 6.2 7.5
RAL 300 103.0 93.0 – 112.3 104.4 93.0 – 115.0 3.9 5.4 6.7
2000 95.0 94.0 – 96.0 97.3 93.5 – 108.0 4.0 2.6 4.8

30 105.4 84.0 – 119.3 104.8 84.0 – 119.3 6.8 8.6 11.0
M8 300 103.9 86.0 – 119.7 103.3 86.0 – 119.7 6.3 7.4 9.7
2000 91.8 80.5 – 112.0 95.8 80.5 – 112.0 9.2 3.7 9.9

300 102.8 94.3 – 111.7 106.1 94.0 – 116.0 5.3 3.4 6.3
APV 3000 105.6 100.3 – 110.0 104.7 94.3 – 110.0 2.9 2.2 3.6
20000 89.9 86.5 – 93.0 92.3 84.0 – 104.0 3.9 0 3.9

300 96.5 88.7 – 108.7 101.0 88.7 – 108.7 4.4 2.3 4.9
DRV 3000 100.1 92.7 – 105.3 99.7 92.7 – 105.3 2.4 0 2.4
20000 85.5 81.0 – 94.5 89.2 81.0 – 100.0 2.8 4.3 5.1

30 105.8 102.7 – 109.7 107.1 93.7 – 116.0 4.5 2.2 5.0
NFV 300 101.2 94.0 – 107.7 106.3 94.0 – 118.7 3.9 4.0 5.6
2000 91.0 83.5 – 100.0 94.9 83.5 – 107.5 5.3 5.4 7.5

30 95.5 86.7 – 109.7 103.0 86.7 – 119.0 8.1 3.6 8.9
SQV 300 99.4 93.7 – 101.7 105.5 92.0 – 117.0 6.0 3.4 6.9
2000 101.8 90.0 – 103.5 99.0 90.0 – 114.5 6.9 0 6.9

30 107.0 101.3 – 114.7 110.3 101.3 – 119.3 3.2 3.7 4.9
ATV 300 103.9 100 – 110.7 107.0 100.0 – 116.7 3.1 3.7 4.8
2000 89.5 84.5 – 93.5 91.9 84.5 – 101.0 2.7 2.0 3.4

30 102.1 97.3 – 105.0 104.4 94.7 – 115.3 3.5 4.8 5.9
RTV 300 105.2 98.0 – 118.7 107.9 98.0 – 118.7 3.9 0 3.9
2000 93.9 87.0 – 102.0 97.5 87.0 – 103.5 4.2 2.5 4.9

30 107.3 100.7 – 112.3 107.7 99.3 – 117.7 3.7 1.3 3.9
LPV 300 108.8 106.3 – 113.3 106.5 101.3 – 113.3 2.3 1.3 2.7
2000 98.8 90.5 – 103.0 99.0 90.5 – 103.0 3.0 0.3 3.0

500 97.3 96.0 – 99.3 97.7 91.3 – 106.0 3.3 1.6 3.7
d4T 7500 100.0 98.7 – 103.2 100.0 92.9 – 105.3 2.6 2.0 3.3
20000 93.5 87.0 – 96.5 95.8 87.0 – 101.5 3.0 1.8 3.5

300 102.2 96.0 – 108.3 102.0 91.3 – 110.0 4.1 0.9 4.2
AZT 3000 104.7 101.0 – 108.0 103.0 95.0 – 112.3 3.6 2.2 4.2
20000 95.5 93.0 – 97.5 97.8 87.5 – 109.0 3.2 4.4 5.4

500 95.8 92.0 – 100.0 98.2 92.0 – 101.3 1.6 1.4 2.2
AZT-G 7500 102.9 96.9 – 107.9 107.8 96.9 – 117.5 4.0 1.6 4.3
20000 108.9 97.5 – 116.5 112.9 93.0 – 118.0 6.9 0 6.9

300 98.4 97.0 – 102.3 103.4 97.0 – 111.0 3.0 2.7 4.0
EFV 3000 103.5 101.7 – 106.0 104.6 99.0 – 111.7 2.1 2.1 3.0
20000 95.8 90.0 – 101.5 93.7 85.5 – 101.5 3.6 1.1 3.8
a

Since accuracy fell outside 80–120% target concentrations for ABC-G and ABC-C QC concentrations, these analytes are considered semiquantitative.