Table V.

Wire landmarks: comparison of differences between two-dimensional and three-dimensional measurements of 20-mm horizontal wires and 40-mm vertical wires for 28 skulls imaged in 3 positions

Wire
orientation
Skull
side
Measurement Difference* Absolute value of
difference


2D 3D 2D 3D
Horizontal Left Mean error (mm) −0.28 0.18 0.5 0.34
% error 1.4% 0.9% 2.5% 1.7%
SD (0.58) (0.38) (0.39) (0.23)
Right Mean error (mm) −0.20 0.03 0.47 0.19
% error 1.0% 0.2% 2.4% 1.0%
SD (0.56) (0.24) (0.34) (0.14)
Vertical Left Mean error (mm) −0.47 −0.23 0.56 0.25
% error 1.2% 0.6% 1.4% 0.6%
SD (0.50) (0.23) (0.39) (0.19)
Right Mean error (mm) −0.50 −0.36 0.61 0.37
% error 1.3% 0.9% 1.5% 0.9%
SD (0.60) (0.26) (0.47) (0.23)

2D, two-dimensional; 3D, three-dimensional.

*

Average of 84 measurements (measurement in image–known length).

Average of 84 measurements |measurement in image–known length|.