Skip to main content
. 2013 May;51(5):1474–1477. doi: 10.1128/JCM.00176-13

Table 1.

Performance of the illumigene group A Streptococcus assay compared to that of the standard culture

Clinical test site no. No. of specimens tested Results (no.)a
Performance (% [95% CI])b
TP FP TN FN Sensitivity Specificity PPV NPV
1 338 40 11 287 0 100 (91–100) 96.3 (93–98) 78.4 (64–88) 100 (98–100)
2 241 18 18 205 0 100 (81–100) 91.9 (87–95) 50 (32–67) 100 (98–100)
3 217 16 13 188 0 100 (79–100) 93.5 (89–96) 55.2 (35–73) 100 (98–100)
Total 796 74 42 680 0 100 (95–100) 94.2 (92–95) 63.8 (54–72) 100 (99–100)
a

TP, true positive; FP, false positive; TN, true negative; FN, false negative.

b

CI, confidence interval; PPV, positive predictive value; NPV, negative predictive value.