Skip to main content
. Author manuscript; available in PMC: 2013 Sep 1.
Published in final edited form as: AJR Am J Roentgenol. 2012 Sep;199(3):W392–W401. doi: 10.2214/AJR.11.7255

Table 2.

Sensitivity and False Positive Marks for the Common Cases

iCAD R2
Sensitivity DG 0.74(119/161)
95% CI: (0.66,0.81)
0.74(119/161)
95% CI: (0.66,0.81)
SF 0.69(111/161)
95% CI: (0.61,0.76)
0.60(97/161)
95% CI: (0.52,0.68)
DGvSF 0.05
95% CI: (−0.03,0.13)
p=0.2559
0.14
95% CI: (0.05,0.23)
p=0.0026
False Positive DG 2.57 ± 1.92 2.07 ± 1.57
SF 3.06 ± 1.72 1.52 ± 1.45
DGvSF −0.49
95% CI: (−0.77,−0.21)p=0.0008
0.55
95% CI: ( 0.33, 0.78)
p<0.0001