Skip to main content
. 2013 May 1;6:9. doi: 10.1186/1755-1536-6-9

Table 1.

Technical evaluation of the BGM competitive ELISA

Technical validation step BGM ELISA
Target
MMP-9 and -12 degradation of biglycan
Lower limit of detection (LDL)
1.54 ng/mL
Dilution recovery of human serum
100% (1:2), 103% (1:4)
Dilution recovery of human plasma
97% (1:2), 99% (1:4)
Dilution recovery of rat serum
107% (1:2), 102% (1:4)
Dilution recovery of rat plasma
89% (1:2)
Dilution recovery of mouse serum
105% (1:2)
Dilution recovery of mouse plasma
97% (1:2)
Intra-assay variation
10%
Inter-assay variation 15%