Table 1.

Risk Probability Number (RPN) Scores for Selected “Top-Ranked” Failure Modes

Failure Mode RPN
Incorrectly exporting films to record and verify system 160
Needing to re-mark patient after treatment shifts are made 160
Incorrect beams and energy are chosen 160
Inability to verify patient’s marks are valid for treatment 128
Incorrect treatment plan generated because of incorrect physician contours 100
Treatment plan and film are exported incorrectly to treatment machine 96
Simulation therapist creates incorrect patient file 96
Radiation therapist write patient setup information on paper leading to incorrect information 96
Therapists incorrectly re-mark patient because of shifts that were written incorrectly on paper 96
Incorrect verification by radiation therapists of paper setup instructions written at the time of simulation 80
Incorrect treatment plan generated because of incorrect beam placement by dosimetry 80
Dosimetrists export dose information to dose calculation program incorrectly 80
Radiation therapist shifts the patient’s position based on treatment plan incorrectly 80
Incorrect instructions on contrast administered to patient 75
Immobilization for patient is incorrect 75