Incorrectly exporting films to record and verify system |
160 |
Needing to re-mark patient after treatment shifts are made |
160 |
Incorrect beams and energy are chosen |
160 |
Inability to verify patient’s marks are valid for treatment |
128 |
Incorrect treatment plan generated because of incorrect physician contours |
100 |
Treatment plan and film are exported incorrectly to treatment machine |
96 |
Simulation therapist creates incorrect patient file |
96 |
Radiation therapist write patient setup information on paper leading to incorrect information |
96 |
Therapists incorrectly re-mark patient because of shifts that were written incorrectly on paper |
96 |
Incorrect verification by radiation therapists of paper setup instructions written at the time of simulation |
80 |
Incorrect treatment plan generated because of incorrect beam placement by dosimetry |
80 |
Dosimetrists export dose information to dose calculation program incorrectly |
80 |
Radiation therapist shifts the patient’s position based on treatment plan incorrectly |
80 |
Incorrect instructions on contrast administered to patient |
75 |
Immobilization for patient is incorrect |
75 |