Skip to main content
. 2013 May 16;8(5):e63968. doi: 10.1371/journal.pone.0063968

Table 1. Comparison of Cq values (Mean ±1 SD; n = 9) from single and duplex real-time PCR assays obtained with dilution series of Borrelia DNA or a synthetic Borrelia 16S oligonucleotide in a background matrix of human genomic DNA.

Copies 16S Synthetic Borrelia 16S Total Borrelia DNA
16S 16S/RNAseP 16S 16S/RNAseP
10,000 26.37±0.28 26.36±0.20 26.44±0.11 26.46±0.11
1,000 29.78±0.22 29.74±0.19 29.89±0.12 29.87±0.06
100 33.17±0.34 33.32±0.24 33.42±0.32 33.12±0.20
10 36.58±0.46 37.03±1.23 37.00±0.77 36.58±0.55
5 37.30±0.72 37.83±1.16 38.33±1.17 37.74±0.85
2.5 38.84±1.25 38.37±0.84 39.02±1.01 38.45±0.84
1 39.49±0.80 39.10±0.98 39.66±0.58 39.52±0.63
PCR Efficiency 104% ±4% 104% ±6% 98% ±1% 103% ±3%
R2 0.99±0.004 0.98±0.01 0.98±0.15 0.99±0.004
Y-Intercept 39.45±0.24 39.48±0.27 40.03±0.19 39.06±1.14
Slope −3.23±0.08 −3.22±0.13 −3.37±0.12 −3.25±0.06