Table 1.
HIFU field parameters in a soft layer placed in the focal region behind the ribs. The array elements located in the shadow of ribs were switched off.
Initial intensity I0, W·cm−2 | 2.5* | 2.5 | 5 | 10 | 20 | 30 | 40 |
Initial pressure p0, MPa | 0.27 | 0.27 | 0.39 | 0.55 | 0.77 | 0.95 | 1.1 |
Focusing gain, GI = IF/I0 | 367 | 377 | 387 | 410 | 476 | 552 | 580 |
Ratio of intensities in the main and side foci Imain/Iside | 1.90 | 1.94 | 1.99 | 2.09 | 2.39 | 2.75 | 2.86 |
Focusing gain, Gp = pF/p0 | 19.2 | 23.5 | 26.2 | 31.8 | 50.9 | 77.1 | 80.8 |
Ratio of pressures in the main and side foci pmain/pside | 1.4 | 1.5 | 1.6 | 1.8 | 2.5 | 3.5 | 3.3 |
Ratio of linear and nonlinear heating rates in the main focus QN/QL | 1.00 | 1.07 | 1.15 | 1.37 | 2.39 | 34.8 | 66.2 |
Ratio of heating rates in the main and side foci Qmain/Qside | 1.90 | 2.00 | 2.12 | 2.42 | 3.92 | 45.4 | 71.3 |
Time of achieving 100°C in the main focus, ms | 19100 | 19000 | 3200 | 740 | 190 | 9.5 | 2.5 |
Ratio of temperature rise in the main and side foci when temperature in the main focus reaches 100 °C, ΔTmain/ΔTside | 1.79 | 1.82 | 2.00 | 2.20 | 3.00 | 26.0 | 65.0 |
The second column, marked as 2.5*W/cm2 for the initial intensity, corresponds to linear acoustic modelling, other initial values correspond to nonlinear modelling.