Table 2. List of complications without revision of the short-stem.
Patient | Implantation | Complication | Time | Notes |
---|---|---|---|---|
♀70 | 09/2005 | Femoral nerve lesion | Intra-op | Completely healed, HHS = 94; follow-up ex. at 4.2 years |
♀59 | 11/2004 | Stem fissure, left without cerclage | Intra-op | Healed without complication, HHS = 100; follow-up ex. at 6 years |
♀55 | 01/2005 | Ceramic head breakage | 12 months | After a considerable fall, replacement of adapter and head, HHS = 96; follow-up ex. at 5 years |
♀69 | 09/2005 | Replacement of cup | 9 days | Without subsequent complications, HHS = 96; follow-up ex. at 4.2 years |
♀57 | 09/2004 | Femoral fissure | Intra-op | Cerclage, without post-op complications, HHS = 100; follow-up ex. at 6.5 years |
♂66 | 01/2005 | Very deep neck resection | Intra-op | No abnormalities, well-integrated, HHS = 100; follow-up ex. at 3.3 years |
♂68 | 04/2005 | Stem subsidence of 9 mm post-op; stable after 1-year follow-up ex. | Post-op | No abnormalities, well-integrated, HHS = 100; follow-up ex. at 5.8 years |
HHS, Harris Hip Score.