| i. Standard (sequential) method only | |
| ■ Initial frequency: | 106 Hz |
| ■ Final frequency: | as indicated in Table 1 |
| ■ Measurement duration: | 0.3 s integration time or minimum of 3 cycles |
| ii. Standard method + Parallel method | |
| ◦ Standard method portion | |
| ■ Initial frequency: | 106 Hz |
| ■ Final frequency: | 10 Hz |
| ■ Measurement duration: | 0.3 s integration time or minimum of 3 cycles |
| ◦ Parallel method portion | |
| ■ Initial frequency: | 10 Hz |
| ■ Final frequency: | as indicated in Table 1 |
| ■ Measurement duration: | 20 s integration time or minimum of 1 cycle |