Skip to main content
. Author manuscript; available in PMC: 2014 Jul 1.
Published in final edited form as: J Pharm Biomed Anal. 2013 Apr 9;0:89–98. doi: 10.1016/j.jpba.2013.03.025

Table 5.

Assay performance data of QCLL, QCL, QCM and QCH belinostat and metabolite concentrations in citrate plasma quantitated against a heparinized plasma calibration curve.

Analyte Concentration (ng/mL) Accuracy (%) Precision (%)
Belinostat 30 (QCLL) −11.4 11.3
100 (QCL) −9.2 5.2
2500 (QCM) 10.5 2.7
4000 (QCH) 12.5 1.4

Belinostat-glucuronide 30 (QCLL) −17.7 6.2
100 (QCL) −9.9 6.6
2500 (QCM) −7.3 5.4
4000 (QCH) −11.5 3.4

Methyl-belinostat 30 (QCLL) −14.7 9.7
100 (QCL) −9.3 5.1
2500 (QCM) −4.9 6.5
4000 (QCH) −9.7 3.9

M21 30 (QCLL) −1.7 9.1
100 (QCL) −1.4 11.1
2500 (QCM) 2.7 5.4
4000 (QCH) 4.1 6.5

M24 30 (QCLL) −2.3 13.7
100 (QCL) −5.5 5.7
2500 (QCM) 10.1 6.5
4000 (QCH) 5.8 6.8

M26 30 (QCLL) 8.7 6.9
100 (QCL) 8.2 3.9
2500 (QCM) 10.0 2.4
4000 (QCH) 8.1 4.7

N=6.