Table 1. Performance characteristics of a multiplexed MRM-MS assay.

Isotope labeled apoA-I was used as the internal standard for all six proteins in the assay and correlation with immunoassay was determined along with linearity and imprecision.

Imprecision
Correlation with
Immunoassay
(N=30)
Linearity LC-MS
(%CV)
Digestion
(%CV)
Overall
(%CV)
A-I 0.95 0.9996 1.15% 0.25% 1.18%
B 0.61 0.9997 4.59% 4.43% 6.38%
C-II 0.92 0.9948 8.38% 7.99% 11.58%
C-III 0.88 0.9976 8.39% 4.41% 9.48%
E 0.92 0.9980 6.03% 4.93% 7.79%
J 0.79 0.9996 9.51% 7.43% 12.07%