. 2012 Nov;146(3):251–260. doi: 10.1111/j.1399-3054.2012.01639.x

Table 1.

Major measurement characteristics of Dualex 4 Scientific. Repeatability (instrumental variations) was evaluated by the standard deviation (sd) and percent standard deviation (%sd) for 30 consecutive measurements on a leaf with mean Chl value of 21 µg cm−2 and mean Flav absorbance of 1.8. Reproducibility (inter-instrument agreement) was obtained from the measurements among five different Dx4 on 80 leaves of four different species. Accuracy for Chl was estimated from the calibration against Chl extracts: root mean square error (RMSE) from Fig. 3 (N = 195) and percent RMSE (%RMSE) for a mean Chl value of 32 µg cm−2. Accuracy for the Flav index was estimated from the comparison to Dualex 3 in absorbance units: RMSE from Fig. 2 (N = 74) and %RMSE for a mean Flav absorbance equal to 1.2

Chlorophyll Epidermal Flavonoids NBI

Source of variation sd %sd sd %sd sd %sd
Repeatability
Clip closed 0.034 0.16 0.004 0.22 0.037 0.31
Clip opened between 0.132 0.62 0.004 0.22 0.063 0.53
RMSE %RMSE RMSE %RMSE

Reproducibility 0.713 2.4 0.034 3.4
Accuracy 5.03 16 0.185 15