. 2013 Apr;51(4):1124–1129. doi: 10.1128/JCM.03113-12

Table 3.

Clinical sensitivity and specificity of the FLULUM assay for 259 specimens submitted in the 2010-2011 seasona

Target Sensitivity
Specificity
TP/(TP+FN) % TN/(TN+FP) %
FluA matrix 146/150 97.3 109/109 100.0
2009 H1 77/85 90.5 174/174 100.0
H3 63/65 96.9 194/194 100.0
FluB matrix 8/9 88.9 250/250 100.0
a

The assay was performed on the Luminex LX200 instrument. Abbreviations: TP, true positive; FN, false negative; TN, true negative; FP,false positive.