Table 2.

Air dose values and DAP values from different cone beam CT units

Dental office Model Tube potential (kV) Current (mA) Exposure time(s) Measured length (cm) Area (cm2) Air dose (mGy)
DAP (mGy cm2)
Mean Standard deviation
1 A 90 5 9 6.7 × 6.8 45.6 2.78 0.10 126.7
2 A 90 4 17.5 6.9 × 6.5 44.9 3.58 0.27 160.6
3 A 90 5 9 6.9 × 6.8 46.9 4.79 0.45 224.7
4 A 90 5 17.5 6.8 × 6.8 46.2 5.87 0.32 271.4
5 A 90 5 17.5 7.0 × 6.8 47.6 7.55 0.31 359.4
6 B 70 10 10.8 6.0 × 4.9 29.4 10.08 0.28 296.4
7 B 74 10 10.8 5.9 × 4.8 28.3 9.19 0.67 260.3
8 B 70 10 10.8 6.2 × 5.0 31.0 9.11 0.43 282.4
9 C 80 5 9.4 6.8 × 6.5 44.2 6.63 0.33 293.0
10 C 80 5 9.4 7.0 × 6.6 46.2 6.43 0.99 297.1
11 C 80 5 9.4 6.9 × 6.5 44.9 7.28 0.89 326.5
12 C 80 5 9.4 6.9 × 6.6 45.5 7.17 0.88 326.5
13 C 80 5 9.4 7.0 × 6.6 46.2 7.61 0.20 351.6
14 D 84 12 12 8.1 × 8.1 65.6 6.14 0.36 402.8
15 D 84 12 12 8.4 × 8.7 73.1 5.67 0.65 414.4
16 D 84 12 12 8.0 × 8.3 66.4 6.46 1.04 428.9
17 D 84 12 12 8.1 × 8.2 66.4 6.64 0.46 441.0
18 D 84 12 12 8.2 × 9.4 77.1 6.49 0.54 500.2
19 E 85 4 17 9.7 × 9.6 93.1 10.33 0.56 961.9
20 E 85 6 17 9.5 × 9.8 93.1 13.11 0.99 1220.5
21 E 85 6 17 9.7 × 9.8 95.1 15.53 0.56 1476.9

DAP, dose–area product.

Models: A, 3D Accuitomo (J Morita, Kyoto, Japan); B, Kodak 9000 3D (Trophy, Croissy-Beaubourg, France); C, Veraviewepocs3D (J Morita); D, ProMax 3D (Planmeca OY, Helsinki, Finland); E, AZ3000 (Asahi Roentgen Industry, Kyoto, Japan).