Table 4.
Layer 1 (dB/cm-MHz) | Layer 2 (dB/cm-MHz) | Layer 3 (dB/cm-MHz) | |||||
---|---|---|---|---|---|---|---|
|
|||||||
Clinical System | Transducer | Const-α | Const-BSC | Const-α | Const-BSC | Const-α | Const-BSC |
Lab characterization | 0.52 | 0.48 | 0.54 | 0.73 | 0.52 | 0.47 | |
UltraSonix RP | L14-5/38 | 0.46 | 0.46 | 0.48 | 0.64 | 0.31 | 0.41 |
Siemens Acuson S2000 | 18L6 | 0.48 | 0.48 | 0.53 | 0.71 | 0.44 | 0.43 |
9L4 | 0.49 | 0.5 | 0.52 | 0.7 | 0.55 | 0.45 | |
VisualSonics Vevo 2100 | MS200 | 0.67 | 0.46 | 0.55 | 0.69 | 0.53 | 0.5 |
Const-α = constant attenuation; Const-BSC = constant backscatter. The mean value was obtained by averaging estimates from different alpha-estimation blocks and from uncorrelated image frames for each layer. For the UltraSonix and Siemens systems, the transducers were placed in contact with Layer 1 and then directed to the more distal layers; for the high-frequency VisualSonics, the transducer was in contact with Layer 3, providing a shorter path to the distal layer (see Figure 1).