Skip to main content
. Author manuscript; available in PMC: 2013 May 30.
Published in final edited form as: Ultrason Imaging. 2012 Oct;34(4):209–221. doi: 10.1177/0161734612464451

Table 4.

Mean Attenuation Coefficient versus Frequency Slope Values (dB/cm-MHz) Estimated Using the Three Clinical Ultrasound Imaging Systems

Layer 1 (dB/cm-MHz) Layer 2 (dB/cm-MHz) Layer 3 (dB/cm-MHz)

Clinical System Transducer Const-α Const-BSC Const-α Const-BSC Const-α Const-BSC
Lab characterization 0.52 0.48 0.54 0.73 0.52 0.47
UltraSonix RP L14-5/38 0.46 0.46 0.48 0.64 0.31 0.41
Siemens Acuson S2000 18L6 0.48 0.48 0.53 0.71 0.44 0.43
9L4 0.49 0.5 0.52 0.7 0.55 0.45
VisualSonics Vevo 2100 MS200 0.67 0.46 0.55 0.69 0.53 0.5

Const-α = constant attenuation; Const-BSC = constant backscatter. The mean value was obtained by averaging estimates from different alpha-estimation blocks and from uncorrelated image frames for each layer. For the UltraSonix and Siemens systems, the transducers were placed in contact with Layer 1 and then directed to the more distal layers; for the high-frequency VisualSonics, the transducer was in contact with Layer 3, providing a shorter path to the distal layer (see Figure 1).