Table 5.
Proximal Layer (% error) |
Mid Layer (% error) |
Distal Layer (% error) |
|||||
---|---|---|---|---|---|---|---|
Clinical System | Transducer | Const-α | Const-BSC | Const-α | Const-BSC | Const-α | Const-BSC |
UltraSonix RP | L14-5/38 | −11.5 | −4.2 | −11.1 | −12.3 | −40.4 | −12.8 |
Siemens Acuson S2000 |
18L6 | −7.7 | 0 | −1.8 | −2.7 | −15.4 | −8.5 |
9L4 | −5.8 | 4.2 | −3.7 | −4.1 | 5.8 | −4.2 | |
VisualSonics Vevo 2100 |
MS200 | 1.9 | −6.4 | 1.8 | −5.5 | 28.8 | 4.2 |
Const-α = constant attenuation; Const-BSC = constant backscatter. The layers are identified (“proximal,” “mid,” and “distal”) according to their position with respect to the transducer during scanning.