Analyte Spiked concentration (ng/mL) Repeatability (CV %) Accuracy (Error %) Interday Precision (CV %) Interday Accuracy (Error %) Recovery (%)
olanzapine 1 19.76 –12.89 12.35 –12.17 73.26
5 13.19 3.92 8.73 –8.91 88.08
20 2.66 –11.65 3.99 –7.08 84.90
40 2.50 7.81 2.56 8.22 84.38
100 0.46 –1.08 3.87 –1.24 93.83
N-desmethyl-olanzpine 1 14.69 –4.04 6.95 –16.45 73.46
5 8.31 11.76 13.95 –5.66 83.68
20 8.64 –5.73 8.23 6.53 89.13
40 3.27 7.92 3.08 5.83 78.31
100 5.63 –3.86 5.08 –2.23 90.41