. 2013 May 28;9:265–272. doi: 10.2147/VHRM.S44569

Table 7.

Screening and recruitment details for the Omron RS3 (HEM-6130-E) validation

Screening and recruitment Recruitment ranges
mmHg All Antihypertensive medication
Total screened 81 SBP
Total excluded 48 Low <90 0 1
 Complete ranges 32 90–129 11
 Range adjustment 0 Medium 130–160 10 3
 Arrhythmias 7 High 161–180 9 4
 Device failure 0 >180 3
 Poor quality sounds 1 DBP
 Cuff size unavailable 0 Low <40 0 1
 Observer disagreement 0 40–79 11
 Distribution 0 Medium 80–100 12 4
 Other reasons 8 High 101–130 10 3
Total recruited 33 >180 0

Abbreviations: SBP, systolic blood pressure; DBP, diastolic blood pressure.