Skip to main content
. 2013 Jun 4;104(11):2476–2484. doi: 10.1016/j.bpj.2013.04.033

Figure 1.

Figure 1

(A) Relation between AFM deflection sensitivity in wet and dry conditions, including a linear fit. (B) Deviation from linearity of the detector as a function of the measured deflection voltage. Dots are data from three separate measurements (down sampled 40× for clarity) and the line is a 3rd order polynomial fit.