Skip to main content
. 2008 Apr 3;8(4):2331–2412. doi: 10.3390/s8042331

Table 2.

Characterization of a number of reported ion selective sensors based on nonconducting polymers for alkali cations.

Cation Ionophore Slope (mV decade-1) Linear Range (M) Most Important Interfering ions (log Ksel> -2) Ref.

Li+-1 lipophilic crown-4 derivatives 58 10-4-10-1 Na+, K+,NH4+ 125
Li+-2 1,10-Phenanthroline Derivatives 58.7 10-4-10-1 Na+, K+ 128

Na+ 1-methyl-1-vinyl-14-crown-5 55.0 3.16×10-6 -1.0×10-1 K+ 132

K+ styrene/4(-vinyl-benzo-24-crown-8) copolymer 58 1.0×10-6 -1.0×10-1 - 140

Rb+ crown ethers incorporating anthraquinone, benzoquinone, and 1,4-dimethoxybezene 54.7 1.0×10-5 -1.0×10-1 Na+,K+,Mg2+,NH4+,Li+ 145

Cs+ calix[4]arene derivative [25-(3-bromo- proploxy)-5,11,-17,23-tetrakis(tert-butyl)-26,27,28-tris(1- propyloxy) calix[4]arene] 58 1.0×10-6-1.0×10-1 - 153