Skip to main content
. 2008 Apr 3;8(4):2331–2412. doi: 10.3390/s8042331

Table 4.

Characterization of a number of reported ion selective sensors based on nonconducting polymers for third main group cations.

Cation Ionophore Slope (mV decade-1) Linear Range (M) Most Important Interfering ions (log Ksel> -2) Ref.

Al+3 xanthone derivative 20.0 1.0 × 10-6-1.6 × 10-1 Hg2+, Ba2+ 188

Ga3+ chlorogallium(III) 30 10-6 −10-2 - 190

In3+ 1-benzyl-3-methyl-4- benzoyl-5-pyrazolone (PMBP) 18.7 3.2×10-5 - 1.0× 10-1 Ga3+ 192

Tl3+-1 2′-amino-1,3,5′-trioxo- spiro[indane-2,4′(5′H)-3′-cyano-indeno(1 .2-b)]pyran 59 0.1-1.0 × 10-6 Rb+, Cs+ 197
Tl3+-2 N′-dioctylethylene-diamine-N′-disuccinic acid 56 6.4 × 10-7- 10-2 Rb+ 200