Skip to main content
. 2013 Feb 1;15(2):511–522. doi: 10.1208/s12248-013-9458-6

Table I.

Volume Averaged Air Velocity and Pressure Drop Data for the Original and 1/3 Inlet Size Devices at Different Flow Rates

Original device 1/3 Inlet size device
30 L/min 60 L/min 100 L/min 30 L/min 60 L/min 100 L/min
Air velocity (m/s) 6.9 14.1 24.1 11.5 23.5 39.5
Pressure drop (kPa) 0.3 1.0 3.5 1.0 4.2 12.0