Skip to main content
. Author manuscript; available in PMC: 2013 Jun 9.
Published in final edited form as: Clin Chem. 2010 Mar 26;56(5):772–780. doi: 10.1373/clinchem.2009.140459

Table 1.

Recovery and imprecision of the assay.

Accuracy
Imprecision, % CV
Amount of Hgb, mg
(relative amount)
Measured enzyme
activitya
Recovery, % Intraassay Interassay
0.56 (100%b) 21.8 100b 2.1 4.5

0.14 (25%) 5.3 97.2 2.5 6.7

0.28 (5%) 1.03 94.8% 4.6 8.2

0.00112 (0.2%) 0.04 92.1% 9.7 13.2
a

Unit is µmol · (g Hgb)−1 · h−1

b

100% was arbitrarily assigned.