Table 1.
Summary of CT hardware and parameters for state-of-the-art CT scanners from the major CT manufacturers
Scanner | X-ray sources, number | Detector rows, number | Detector-row x-axis dimension, mm* | Total nominal beam width, mm | Fastest gantry rotation time, s | Temporal resolution for each cross-sectional image, s† |
---|---|---|---|---|---|---|
GE Discovery CT750 HD |
1 | 64 | 0.625 | 40 | 0.35 | 0.175 |
Hitachi SCENARIA | 1 | 64 | 0.625 | 40 | 0.35 | 0.175 |
Phillips Brilliance iCT | 1 | 128‡ | 0.625 | 80 | 0.27 | 0.135 |
Siemens SOMATOM Definition FLASH |
2 (95° apart) | 64‡ | 0.6 | 40 | 0.28 | 0.075 |
Toshiba Aquilion ONE | 1 | 320 | 0.5 | 160 | 0.35 | 0.175 |
Values measured at scanner isocenter;
Values do not reflect the use of multi-segment reconstruction;
Uses z-axis flying focal spot to sample each detector twice per rotation.