Table 1.

Summary of CT hardware and parameters for state-of-the-art CT scanners from the major CT manufacturers

Scanner X-ray sources, number Detector rows, number Detector-row x-axis dimension, mm* Total nominal beam width, mm Fastest gantry rotation time, s Temporal resolution for each cross-sectional image, s
GE Discovery CT750
HD
1 64 0.625 40 0.35 0.175
Hitachi SCENARIA 1 64 0.625 40 0.35 0.175
Phillips Brilliance iCT 1 128 0.625 80 0.27 0.135
Siemens SOMATOM
Definition FLASH
2 (95° apart) 64 0.6 40 0.28 0.075
Toshiba Aquilion ONE 1 320 0.5 160 0.35 0.175
*

Values measured at scanner isocenter;

Values do not reflect the use of multi-segment reconstruction;

Uses z-axis flying focal spot to sample each detector twice per rotation.