Table 2.

Heart rate thresholds for scan mode selection for coronary angiography on state-of-the art CT scanners

Scan mode Heart rate threshold, bpm
Prospective ECG-triggered helical ≤60 bpm
Prospective ECG-triggered axial ≤65–80 bpm*
Retrospective ECG-gated helical ≥65–80 bpm*
*

Threshold is higher for scanners with higher temporal resolution.

Bpm, beats per minute; CT, computed tomography; ECG, electrocardiogram.