TABLE 3. Relative Noise Estimation.

Estimates of relative noise between corresponding iBH- and eBH-CT were determined by regional assessment of the distribution of image intensities within manually placed ROIs. The ratio of COVs was calculated from these measurements as a surrogate for the relative noise between image pairs. On average, the estimated eBH-CT noise is 2.2× that of the corresponding iBH-CT.

Case
Number
Inhale COV Exhale COV Ratio (Exhale /
Inhale)
1 0.011 0.018 1.64
2 0.028 0.042 1.50
3 0.015 0.031 2.07
4 0.012 0.024 2.00
5 0.016 0.031 1.94
6 0.008 0.019 2.37
7 0.014 0.037 2.64
8 0.012 0.026 2.17
9 0.014 0.039 2.78
10 0.013 0.037 2.85

Average Ratio 2.20