Table 2. Baseline measured physiologic parameters for ventilation with the SAVe I, SAVe II, or Siemens Servo 900C.
SAVe I(n=6) | SAVe II (n=6) | |||
---|---|---|---|---|
Variable | Servo | SAVe I | Servo | SAVe II |
PaO2, torr | 92.5 ± 9.1 | 92.5 ± 14.1 | 96.3 ± 3.7 | 98.5 ± 6.3 |
PaCO2, torr | 37.7 ± 1.8* | 27.7 ± 4.2* | 38.3 ± 2.9 | 36.5 ± 3.1 |
VE, L/min | 5.7 ± 1.5† | 8.3 ± 0.8† | 5.3 ± 1.6 | 5.1 ± 1.3 |
Peak airway
pressure, cm H2O |
16.7 ± 2.1 | 14.7 ± 3.4 | 16.9 ± 2.3§ | 20.5 ± 2.5§ |
Compliance,
ml/cm H2O |
21.3 ± 9.0 | 15.6 ± 5.2 | 27.8 ± 10.4‡ | 16.8 ± 3.0‡ |
PaO2/FIO2 ratio | 440.5 ± 43.2 | 440.5 ± 67.1 | 458.7 ± 17.5 | 463.9 ± 24.0 |
Mean BP, mm
Hg |
106.5 ± 9.8 | 99.8 ± 13.9 | 97.7 ± 10.2 | 97.7 ± 7.3 |
Mean
pulmonary artery pressure, mm Hg |
17.5 ± 3.0 | 15.6 ± 3.3 | 15.6 ± 2.6 | 15.9 ± 2.3 |
Cardiac
Output, L/min |
2.47 ± 0.40 | 2.83 ± 0.60 | 2.58 ± 0.62 | 2.79 ± 0.57 |
PAOP | 7.63 ± 0.98+ | 5.83 ± 0.98+ | 7.50 ± 1.87 | 7.42 ± 2.11 |
Values are mean ± SD. Comparisons are made between Siemens Servo and SAVe ventilator in each group. p>0.05 unless specified.
p=0.001
p=0.007
p=0.02
p=0.048
p=0.01.