Table 3. Physiologic variables at failure of mechanical ventilation or peak lung injury while being supported with SAVe I or Siemens 900C. (n=6).
Variable | Servo | SAVe I |
---|---|---|
PaO2, torr | 106.0 ± 25.6* | 52.0 ± 11.1* |
PaCO2, torr | 57.8 ± 11.3 | 56.3 ± 19.7 |
VE, L/min | 5.9 ± 2.0 | 5.5 ± 1.1 |
Compliance, ml/cm H2O | 8.8 ± 1.9 | 5.9 ± 2.9 |
PaO2/FIO2 ratio | 147.2 ± 71.1 | 93.2 ± 33.2 |
Mean BP, mm Hg | 97.3 ± 18.4 | 106.7 ± 25.9 |
Mean pulmonary artery
pressure, mm Hg |
35.1 ± 9.1 | 46.7 ± 17.2 |
Cardiac Output, L/min | 1.55 ± 0.53 | 1.71 ± 0.51 |
PAOP, mm Hg | 12.50 ± 4.04 | 14.30 ± 6.06 |
PEEP, cm H2O | 6.7 ± 3.2† | 0† |
FIO2 | 0.80 ± 0.25 | 0.58 ± 0.06 |
Peak airway pressure,
cm H2O |
30.8 ± 7.2 | 29.5 ± 6.9 |
Values are mean ± SD. Comparisons are made between Siemens Servo and SAVe I ventilator. p>0.05 unless specified.
p=0.002
p=0.003.