Table 4. Physiologic variables at failure of mechanical ventilation or peak lung injury while being supported with SAVe II or Siemens 900C. (n=6).
Variable | Servo | SAVe II |
---|---|---|
PaO2, torr | 158.3 ± 167.7 | 141.8 ± 169.3 |
PaCO2, torr | 58.5 ± 17.0 | 52.5 ± 10.4 |
VE, L/min | 5.2 ± 1.4 | 5.0 ± 1.3 |
Compliance, ml/cm H2O | 10.4 ± 4.1 | 8.6 ± 3.8 |
PaO2/FIO2 ratio | 244.2 ± 140.8 | 224.8 ± 145.8 |
Mean BP, mm Hg | 94.5 ± 22.5 | 90.8 ± 22.0 |
Mean pulmonary artery
pressure, mm Hg |
32.2 ± 5.4 | 32.7 ± 4.0 |
Cardiac Output, L/min | 2.04 ± 0.72 | 2.03 ± 0.75 |
PAOP, mm Hg | 10.58 ± 3.77 | 11.50 ± 4.85 |
PEEP, cm H2O | 7.7 ± 2.6 | 5.7 ± 1.2 |
FIO2 | 0.61 ± 0.30 | 0.58 ± 0.30 |
Peak airway pressure,
cm H2O |
31.7 ± 7.3 | 36.2 ± 10.8 |
Values are mean ± SD. Comparisons are made between Siemens Servo and SAVe II ventilator. p>0.05 for all comparisons.